Microfabricated multiple field of view imaging flow cytometry.

Citation:

Schonbrun E, Gorthi SS, Schaak D. Microfabricated multiple field of view imaging flow cytometry. Lab Chip 2012;12(2):268-73.

Abstract:

The combination of microscopy and flow cytometry enables image based screening of large collections of cells. Despite the proposition more than thirty years ago, adding high resolution wide-field imaging to flow cytometers remains challenging. The velocity of cells in flow cytometry can surpass a meter per second, requiring either sub-microsecond exposure times or other sophisticated photodetection techniques. Instead of faster detectors and brighter sources, we demonstrate that by imaging multiple channels simultaneously, a high throughput can be maintained with a flow velocity reduced in proportion to the degree of parallelization. The multi-field of view imaging flow cytometer (MIFC) is implemented with parallel arrays of microfluidic channels and diffractive lenses that produce sixteen wide field images with a magnification of 45 and submicron resolution. Using this device, we have imaged latex beads, red blood cells, and acute myeloid leukemia cells at rates of 2,000-20,000 per second.